France

GlucoFix Tech

   
   

GlucoMen Areo - Menarini Diagnostics

MAIN FEATURES

   

GlucoMen Areo - Menarini Diagnostics

NFC

   

GlucoMen Areo - Menarini Diagnostics

TECHNICAL DATA

GlucoFix Tech

Available in: France, Netherlands

   

MAIN FEATURES

NO CODING

Maximum accuracy and no risk of error from incorrect coding.

 

GlucoMen Areo - Menarini Diagnostics

MARKERS

Three markers for before and after meals and exercise.

 

GlucoMen Areo - Menarini Diagnostics

DIABETES REMOTE CONTROL

Send your data by email or text message.

 

GlucoMen Areo - Menarini Diagnostics

NFC

Transfer the data from your GLUCOFIX® TECH onto your smartphone ortablet simply...

by bringing them close together!

 
     

NCF - Glucomen Areo - Menarini Diagnostics

NFC (Near Field Communication) is an innovative data-transmission system that makes it possible to transfer
the blood glucose results of your GLUCOFIX® TECH onto a smartphone and tablet simply... by bringing them
close together!

Downloading your blood sugar data onto your smartphone and tablet with the NFC system is quick and easy!
Once you have installed the dedicated A. Menarini Diagnostics app you will be able to:

GlucoLog Lite App - Menarini Diagnostics

Instantly view the graphs and statistics of your blood sugar data with the GlucoLog® Lite app
and send them by email to your physician or a person of your trust.

   

GlucoLog Mobile - app - Menarini Diagnostics

Send and share your results in real time by text message with the GlucoLog® Mobile app.


   

TECHNICAL DATA

 

Test strip:

GLUCOFIX® TECH Sensor

Assay Method:

Electrochemical, Glucose Oxidase

Icodextrin interference GLUCOFIX TECH system is suitable for use by people on peritoneal dialysis using solutions containing icodextrin.

Test Range:

20 – 600 mg/dL (1.1 – 33.3 mmol/L)

Haematocrit range:

10 – 70 %

Sample size:

Minimum 0.5 μL

Test time:

5 seconds

Batteries:

Two 3.0 V lithium batteries (CR2032)

Battery life:

A minimum of 1000 tests or about 1 year

Memory:

730 test results with markers, dates and times

Alarms:

Up to three settable acoustic alarms

Averages:

Over 1, 7, 14, 30, 60, 90 days

Dimensions:

85.5 mm (L) x 56 mm (W) x 18.2 mm (H)

Weight:

46 g (without batteries)

Meter Operating Conditions:

Temperature: 5 - 45 °C / (41 - 113 °F)
Relative Humidity: 20 - 90%

Our values

4 values, which have given life to Menarini and shaped its history

READ MORE »