GLUCOFIX TECH 2K

Available in France and Netherlands
Test strip

GLUCOFIX TECH Sensor

Sample
Fresh capillary whole blood
Sample volume
0.5 µL
Measuring principle
Electrochemical
Enzyme system
Glucose oxidase
Testing time
5 seconds
Test range
20-600 mg/dL (1.1-33.3 mmol/L)
Haematocrit range
10-70% (haematocrit compensated)
Testing conditions
Temperature: 5-45 °C (41-113 °F)
Relative humidity: 20-90% (no dew condensation)
Icodextrin interference
The GLUCOFIX TECH 2K system is suitable for use by
people on peritoneal dialysis using solutions containing
icodextrin
Device standard
EN ISO 15197:2015
Batteries
Two 3.0 V lithium batteries (CR2032)
Battery life
Approx. 900 tests
Memory
730 glucose test results
100 ß-ketone test results
Alarms
Up to 6 settable acoustic alarms
(3 for glucose, 3 for ß-ketone measurements)
Averages
Over 1, 7, 14, 30, 60, 90 days
(for glucose results only)
Dimensions
85 mm x 56 mm x 18 mm
Weight
52 g
Data transmission
To PC (via USB cable)
To mobile devices (via NFC or external Bluetooth)

β-KETONE MEASUREMENT SPECIFICATIONS

Test strip
GLUCOFIX TECH ß-Ketone Sensor
Sample volume
0.8 µL
Measuring principle
Electrochemical
Enzyme system
ß-Hydroxybutyrate dehydrogenase
Testing time
8 seconds
Test range
0.1-8.0 mmol/L
Haematocrit range
20-60% (haematocrit compensated)
Testing conditions
Temperature: 10-40 °C (50-104 °F)
Relative humidity: <85%
Packaging
Each test strip is wrapped individually in an aluminium
foil pouch