France

Specifications

Analysis method

Latex agglutination method

Principles of measurement

Integrated spherical nephelometry (turbidity) determination

Through put

200 samples/hour

Volume of test sample

10 ~ 20μL

Volume of reagent

Latex: 90μL Buffer : 190μL

Sample table

80 test sample ( Rack method )

Test sample tube

Exclusive feces pickers (HemoAuto MC)

Pipette for test sample

Direct insertion into feces pickers

Pipette for reagents

Pressed off nozzle system with buffer solution

Reaction table

Capacity for reaction cells placed: 40 cells

Reaction time

4.5 minutes

Supply of reaction cells

Automatic supply and discharge

Total number of reaction cells

80 cells (two cassettes with 40 reaction cells each)

Reaction temperature

Room temperature

Light source

Halogen lamp 5V 9W

Detector

Silicon photodiode

Calibration of concentration

Seven-point calibration curve.

Calibration curve

Master-curve system, two-point correction

Monitor

Color Liquid-crystal monitor ( Touch panel )

Printer

32-figures thermal printer

Power supply

AC 100V ±10V or AC 220V±10V

Power consumption

300 VA

Sizes

H 500 × W 600 × D 610 mm

Weight

56 kg

External output

RS232C

Our values

4 values, which have given life to Menarini and shaped its history