France

HB9210 PREMIER

The information to be found in this section is intended for the exclusive use of healthcare professionals.

Dimension
74 cm H, 54 cm W, 66 cm D
   
Weight
62 Kg
   
Element of analysis
Quantification of glycated haemoglobin
   
Principle
Boronate affinity high performance liquid chromatography
   
Minimum readable division
Area percent = 0,01%
   
Minimum sample requirements
10 µL whole blood, 5 µL packed red blood cells
   
Dilution
Whole blood 1 : 150, packed red blood cells 1 : 300
   
Sample capacity
210 samples + STAT position
   
Throughput
75 seconds per injection
   
Analytical column
Boronate bonded to a porous polymer support gel
   
Column temperature
55 °C
   
Detection
LED wavelength detector, 413 nm
   
Software
lnstalled on hard drive of Hb9210 Premier CPU
   
Calibrators
On board, Glycated Haemoglobin Calibrators (2 levels)
   
Controls
On board, Glycated Haemoglobin Controls (2 levels)
   

Our values

4 values, which have given life to Menarini and shaped its history